An inverse photoemission spectrometer has been built and tested. It consists of a photon detector and a low energy electron gun in ultrahigh vacuum chamber (about $10^{-10}$ torr). The photon detector which we have bulit is a band pass photon detector of which resolution is about 0.6 eV at mean photon energy $h_ω$ = 9.8 eV. It consists of an electon multiplier with CuBe cathode and a $CaF_2$ optical cutoff window. The electron gun originally designed by Erdman and Zipf has been constructed. The extractor of the electron optics has been modified to enhance electron beam current. The electron trajectory calculated by ray-tracing program for the electron gun is described. The performance of the electron gun is presented. An isochromat spectrum for uncleaned Si(111) surface was obtained.
본 논문에서는 역광전자분광기의 작동을 시험해 보았다. 전체 분해능은 0.86 eV 였다.
전자총은 촛점이 잘 맺혔고, 동작온도는 높았지만 전자의 에너지 10 eV에서 6 ㎂의 전자가 흘렀다. 광검출기는 본 데이타 취득 시스템에서 약 10 카운트/초/㎂ 로 적당한 감도를 보여 주었다.
실리콘 시료에 대해 신호대 잡음비가 약 10 정도의 isochromat 를 얻었다. 이 실리콘의 역광전자분광 시험은 고체 표면의 비점유전자상태를 고찰할 수 있는 가능성을 보여주고 있다.