In this paper, a study on the small size plasma power source and discharge structures for reduced treatment time, proposed new system and discharge structures.
The power source to present, even though able to operate at a high frequency of 27MHz, to generate an output voltage of 10kV or more, in a portable size, Further, it presents the structure of the dual resonant inverter, thereby large resonant current circulating and reduced stress to the switch. Further, it is operated in a pulsed mode within usec can be to minimize the generation of ozone.
The proposed resonant frequency sensing a stable way to enable self-resonant in a short period, complemented through Lead sensing problems with dealy of the switch drive circuit.
By presenting a new discharge structure, and to allow low-power, large-area, uniform discharge.
본 논문에서는 소형 플라즈마 전원장치에 관한 연구와 치료 시간 감소를 위한 방전구조를 연구하여 각각 새로운 시스템과 방전구조를 제시하였다.
제시하는 전원장치는 27MHz 고주파로 동작하면서 10kV 이상의 출력 전압을 발생시킬 수 있으면서도 휴대 가능한 크기이다. 또한 이중 공진 인버터 구조를 제시하여 큰 공진전류 순환시켜 스위치에 부담을 줄였다. 또한 오존의 발생을 최소화하기 위하여 수 usec이내의 펄스모드로 동작한다.
짧은 기간 내에 자체공진이 가능하도록 안정적인 공진주파수 sensing 방법을 제안하였고, 구동회로의 dealy에 의한 문제를 Lead sensing을 통하여 보완하였다.
새로운 방전구조를 제시하여 저전력, 대면적, 균일한 방전이 가능하도록 하였다.