In this thesis, the effects of motion in NMR imaging is studied. A new phase encoding method is proposed to obtain the flow images for high speed velocity. The measurement of high speed velocity as well as the methods of moving artifact correction are discussed according to the motion types which can be classified from spin density variation and the change in flow velocity of spins in the sample. In order to correct moving artifacts, phase compensation using phase coding gradient in pulse sequence is suggested. In addition, as an application to the real human imaging, E.C.G. (electro-cardiogram) gated imaging is performed to obtain blood velocity images of descending aorta and density images of the heart.
Computer simulations as well as experiments are carried out and results are presented.
핵자기 공명 단층 촬영에 있어서 움직이는 물체의 영상에 대한 고찰을 하였다. 특히 새롭게 개선된 위상 부호화 방법에 의해 종전보다 빠른 위속 측정에 관한 이론과 알고리즘을 연구 하였다. 또한 물체가 주기성을 가지고 움직일 경우 올바른 영상을 얻기 위한 한 가지 방법의 예로써 인체의 심장 영상을 얻기 위해 E.C.G. 신호에 펄스 순서를 맞추어 인체 실험을 하였다. 그리고 이것을 응용하여 인체 내의 혈액 순환 속도가 빠르고 그 속도가 주기성을 띤 하대 동맥의 유속 영상을 얻었다.