A study of spin density(ρ) and relaxation time ($T_1$, $T_2$) distributions in NMR imaging is presented. Spin density and spin-lattice relaxation time calculations are performed using two repeated images which differ in repetition times while $T_2$ calculation is performed using the multiple spin-echo method. It is shown that the results obtained by the saturation recovery method with spin-echo(SRSE) is as good as the one obtained by the inversion recovery technique for the measurement of $T_1$ according to the criterion defined by the minimum error multiplied by the square root of imaging time. The optimal pulse timings for SRSE and multiple spin-echo are determined through the computer simulation based on the expected experimental conditions.
The results obtained by computer simulations and experiments using the slice imaging with line-integral projection reconstruction are presented to demonstrate the advantages of SRSE. Similar results are expected for other imaging methods such as the total volume imaging with plane-integral projection reconstruction as well as 2-D and 3-D direct Fourier imaging.
핵자기 공명 단층 촬영에 있어서 핵 밀도와 $T_1$, $T_2$ 영상의 계산에 대해서 연구되었다. 핵밀도와 $T_1$ 은 반복 주기가 서로 다른 두 영상을 이용하였으며 $T_2$의 계산은 multiple spin-echo 방법을 이용하였다. 오차와 영상 획득 시간을 동시에 고려해 볼때, $T_1$의 측정에 있어서 spin-echo 를 쓴 포화 회복 방법이 변전 회복 방법보다 우수함을 보인다. 이를 여러 방법의 최적 펄스 타이밍을 컴퓨터 시뮬레이션으로 보였다. 위의 서술은 선 적분 값을 이용한 방법에 대해서 적용되어 졌으나 다른 방법에 대해서도 똑같이 적용될 것으로 생각된다.