A three-channel surface gradient coil(SGC) which is optimized for high resolution NMR imaging of a large object with good spatial linearity is presented. The SGC provides strong gradient fields as well as large size of available object with existing gradient drivers which are the fundamental requirements of high resolution human body imaging.
Non-linear gradient field of a the SGC is a problem in real application Geometry and current distribution of the SGC are, therefore, optimized to maximize the linearity and strength of the gradient field and minimize inductance and resistance.
To compensate SNR loss of high resolution imaging, a surface RF coil is integrated onto the SGC and thereby devices a thin NMR imaging probe. Geometrical structures and characteristics of the proposed three channel surface gradient coil are discussed and some experimental results as well as computer simulation results are presented. By using the assembly, a high resolution($\triangle{r} \approx 200\mu m) and good SNR image of volunteer's spine was obtained with whole-body KAIS 2.0T MRI system.
자기 공명 영상법에서 고해상도 영상에 필요한 강한 경사자계를 평면구조의 코일을 사용하여 얻었다. 이 평면구조의 코일은 대상 물체의 크기에 제약을 받지않고 강한 경사자계를 얻을수 있는 장점이 있으나, 자계의 비 선형성이 문제점으로 지적되어 왔다.
최적화 과정을 통해 이러한 문제점을 최소화 하였으며, 저항값과 인덕턴스를 실제 사용 가능하도록 줄였다. 또한, 신호대 잡음비를 개선하기위하여 평면형의 RF 코일을 평면 경사 자계코일과 결합시켜 사용했다.
본 논문은 이 평면형 코일의 설계 과정, 구조및 특성에관하여 설명 하였으며, 모의 실험과 200㎛의 해상도를 갖는 척추의 실제 영상을 보였다.