The absolute activities of $^{55}Fe$ and $^{75}Se$ which are electron capture nuclides was determined. The activity of $^{55}Fe$ nuclide which emits only characteristic X-rays and Auger electrons was determined by X-ray spectrometry. The X-ray spectrum of $^{55}Fe$ was measured by Si(Li) detector and the peak area was calculated by summation method. The efficiency of Si(Li) detector was determined by using $^{51}Cr$, $^{54}Mn$ and $^{57}Co$ standard sources, the measured efficiency was corrected for the attenuation effects of the materials between a source and the detector, and for the probability of X-ray escape from detector crystal. The corrected efficiency values have a value of nearly constant which is independent of X-ray energies. Also the efficiency ε' which means ε·$\omega_k$ was used to determine the activity of $^{55}Fe$ in order to eliminate the uncertainty of fluorescence yield. The activity of $^{55}Fe$ was 116.72 MBq/g as of January 1, 1990 0 h (UT) and the combined uncertainty was found to be 0.84 MBq(0.72%).
However, the activity of $^{75}Se$ which is isomeric transition nuclide was measured by 4π(e,x)-τ coincidence counting method. The efficiency of the 4π proportional counter was changed by the variation of low energy threshold level, and the range of the efficiency variation was from 25% to 68%.
The result obtained by efficiency extrapolation method was corrected for the delayed transition effect of meta-stable state of 304 keV level. The delayed transition effect was evaluated from the theoretical calculation of the counting probability of the 4π proportional counter and the result was (5.9±0.47) %. The activity of $^{75}Se$ was 1628.04 kBq/g as of January 1, 1991 0 h(UT) and the combined uncertainty was found to be 7.81 kBq(0.48%). The activity of $^{75}Se$ was also determined by gamma-ray spectrometry, and the activity was intercompared with the activity which was measured by 4π(e,x)-τ coincidence counting method. The activities obtained from the two methods show a difference of 0.41%, so they are in good agreement with each other in the uncertainty range.
전자포획핵종인 $^{55}Fe$와 $^{75}Se$의 방사능을 절대측정 하였다.
특성 X-선과 오제전자만을 방출하는 $^{55}Fe$ 핵종의 방사능은 X-선 분광분석 방법에 의하여 결정되었다.
X-선 스펙트럼은 Si(Li) 반도체 검출기에 의해 측정되었으며 스펙트럼 peak 면적은 summation 방법에 의해 산출하였다. Si(Li) 검출기의 효율은 $^{51}Cr$, $^{54}Mn$, $^{57}Co$ 표준선원을 사용하여 결정되었으며, 산출된 효율값은 검출기와 측정선원 사이의 매질에 대한 감쇠효과와 검출기 결정으로 부터의 X-선 탈출 확률에 대하여 보정하였다. 보정된 효율값은 X-선 에너지에 의존하지 않는 상수값을 보여주고 있다. 또한 X-선 방출확률에 의한 방사능 측정의 불확도를 최소화 하기 위해 ε‘=ε$\omega_k$ 에 의해 산출된 효율값으로 부터 $^{55}Fe$ 핵종의 방사능을 결정하였다. 측정결과는 1990년 1월 1일 0시(UT) 현재 116.72 MBq/g 으로 나타났으며 전체불확도는 0.84 MBq(0.72%)의 좋은 결과를 얻었다.
한편 이성핵 전이 핵종인 $^{75}Se$의 방사능은 4π(e,x)-τ 동시계수방법에 의해 측정되었다. 저에너지 문턱준위 변화 방법에 의해 산출된 4π 비례계수기의 효율변화 범위는 25~68 %로 나타났으며 효율외삽 방법에 의해 방사능을 산출 하였다. 산출된 방사능은 304 keV 준위의 준안정상태에 대한 지발전이 효과에 대하여 보정되었다. 지발전이 효과는 4π 비례계수기의 계수확률에 대한 이론적 계산으로 부터 산출되었으며 그 결과는 (5.9±0.49) %로 나타났다. $^{75}Se$ 핵종의 방사능은 1991년 1월 1일 0시(UT) 현재 1628.04 kBq/g으로 나타났으며 전체 불확도는 7.81 kBq(0.48 %)의 좋은 결과를 얻었다.
또한 4π(e,x)-τ 동시계수 방법에 의해 결정된 $^{75}Se$의 방사능은 감마선 분광 분석방법에 의해 결정된 방사능과 상호비교 되어졌다. 두 가지 방법에 의해 결정된 $^{75}Se$ 핵종의 방사능 차이는 0.41 %로 나타났으며 전체불확도 범위내에서 잘 일치하고 있음을 보여주고 있다.