In conventional KWE method, pulsatile nature of blood flow makes zipperlike artifact along the direction of phase encoding. On the other hand, line-integral projection reconstruction (LPR) method has some advantages such as imaging of short $T_2$ object and elimination of the phase fluctuation due to flow. The LPR, however, necessarily requires filtering and back-projection processes, so that the reconstruction takes a lot of time. To overcome the long reconstruction time of LPR, a concentric square raster sampling scheme (CSRS) method was developed. The CSRS is a fast reconstruction method that has same properties as LPR. In this paper, a new method of flow artifact reduction using CSRS is proposed. Through computer simulations and experiments, it was verified that the proposed technique can eliminate phase fluctuations, reduce flow artifact, and shorten the reconstruction time. All experiments were performed with 2.0 T KAIS NMR system.
핵자기공명 영상법에서 혈류에 의한 영상 훼손이 위상 변환 방향으로 나타난다. 선형 적분 투영법(LPR)은 이러한 영상 훼손을 줄일 수 있을 뿐만 아니라 짧은 T2 성분을 영상화 할 수 있는 장점이 있다. 그러나 선형 적분 투영법으로 영상을 얻기 위해 필요한 filtering과 backprojection 과정은 매우 많은 계산량을 요하므로 긴 처리 시간이 걸린다. 한편 동심 방사 표본화 방식(CSRS)에 의한 영상법은 LPR에서의 장점을 그대로 가지면서 영상을 재구성하는 처리 시간이 짧은 영상법이다. 이 논문에서는 CSRS를 이용하여 혈류에 의한 영상 훼손을 감소시키고 영상처리 시간을 줄이는 방법을 제안하였다. 컴퓨터 simulation과 phantom 실험을 통하여 그 효과를 입증하였으며 이방법을 임상실험에도 적용하였다.