Circular cone-beam computed tomography (CBCT) provides three-dimensional anatomy of an object from the projection data acquired in a single-rotation scan. Due to the large illumination area, a high level of scatter limits the image quality of CBCT by reducing the image contrast, increasing the image noise, and introducing inaccurate CT numbers in the reconstructed images. In this work, a scatter correction method using beam-block strips was presented. A single scan was performed on a cylindrical uniform phantom, and scatter was estimated from each projection data. Using beam-block strips that were positioned in a longitudinal direction to the rotation axis, detected signals in the shadow regions were estimated through interpolation. In addition, a separate estimation of signals was made through interpolation in the illuminated regions; in order to get rid of the effect of unwanted partial transmission through the beam-blocking material. The estimated scatter was extracted from the two estimates. A backprojection filtration (BPF) algorithm was applied to image reconstruction. The results showed that about 30% of the error due to scatter has been corrected by the proposed algorithm.
원형 콘빔 CT는 한번의 스캔으로부터 피사체의 3차원적 해부학적 정보를 줄 수 있다. 한편, 콘빔 CT 는 넓은 영역을 조사하기 때문에 피사체에 의한 산란이 영상 품질을 크게 떨어뜨릴 수 있다. 이는 영상 대조도의 저하, 노이즈의 증가, 및 CT 값의 부정확 등으로 나타난다. 본 연구에서는 빔차단 스트립을 이용한 산란 보정 방법에 대해 연구했다. 원통형의 균일한 팬텀을 이용하여 회전축에 나란하게 빔차단 스트립을 배열하고, 스캔을 한 다음 각 프로젝션에서 빔차단 스트립에 의한 그림자 부분을 내삽하여 산란 분포를 추출하였다. 빔차단 물질을 통하여 부분적으로 투과된 엑스선의 영향을 제거하기 위하여 스트립 사이의 열린 공간에서 얻은 데이터 또한 내삽하여 부분 투과 분포를 앞서 구한 산란 분포에서 차감하였다. 역투영 필터링 방법을 이용하여 영상 재건을 도모하였고, 본 연구에서 제안한 산란 보정의 결과 약 30% 정도의 영상 오차를 줄일 수 있었다.