Recently ultra fast MR imaging using DANTE sequence have been reported. The advantages of the fast imaging using DANTE pulse sequence are that the method does not require high speed gradient switching on the contrary to the other fast imagings techniques such as the Echo Planar Imaging (EPI) and Spiral Scan Eho Planar Imaging (SEPI) techniques where submillisecond gradient switching are the basic requirement. DANTE fast imaging unlike the other gradient switching based techniques it can be implemented on the conventional MRI system.
On the other hand, the original DANTE pulse sequence, excites only a small fraction of imaging voxel is usually selected, that means only a small fraction of the magnetization in each imaging pixel is used, therefore, results in a poor signal to noise ratio (SNR). In this paper, we analyze the SNR of various of DANTE fast imaging sequences and proposed a new FM shaped DANTE sequence with which a substantial amount of the signal to noise ratio can be improved.
최근들어 DANTE 펄스를 이용한 초고속 핵자기 공명 영상법이 개발되고 있다. DANTE 펄스를 이용한 고속 영상법은 기존의 방법인 EPI 와 SEPI 방법보다 더욱 빠른 영상법이고 고속 스위칭 경사자계가 필요없는 RF 스위칭 방법이다. 그래서 더욱 빠른 영상을 할 수 있는 것이다.
DANTE 펄스는 각 영상의 픽셀에서 차지하는 selectivity(선택도)가 적기 때문에 신호 대 잡음비가 매우 나쁘다. 이러한 신호 대 잡음비를 높이기 위해 주파수 변조 방식을 이용한 DANTE 펄스를 소개하고자 한다. 이로 인해 신호 대 잡음이 훨씬 나아진 것을 알 수 있다.