In this thesis, the characteristics of the practical application of the superficial electronic brachytherapy device developed for the treatment of skin cancer were found and measured. The measured characteristics are uniformity, absorbed dose rate at skin cancer site. The size of the fabricated surface applicator was 10 mm, 20 mm to be applicable for various skin cancer sizes. First, uniformity was measured. The uniformity was divided into flatness, symmetry, and penumbra. These values were measured and the measured values were verified to have met standards. Second, absorbed dose rate at skin cancer site was measured. In order to measure that, absorbed dose rate at skin surface and percentage depth dose were obtained. After measuring the characteristics, cell experiment and animal experiment were performed to verify the therapeutic effect. Cell experiments and animal experiments have shown that Superficial EBT devices can be used for practical treatment of skin cancer. In addition, similar results with currently used LINAC equipment will be helpful for future commercialization.
이 논문에서는 피부암 치료를 위해 개발된 소형 피상 전자 근접치료장비의 실제 적용을 위해 필요한 특성들을 찾고 측정하였다. 그 특성들은 크게 균일도, 피부암 위치에서의 흡수선량률이다. 다양한 피부암 크기에 적용하기 위해 크기 10 mm, 20 mm의 표면 어플리케이터가 제작되었다. 먼저 균일도가 측정되었다. 균일도를 크게 편평도, 대칭도, 반영으로 나누어 측정했다. 측정된 모든 값들은 치료를 위해 필요한 값들을 만족했다. 다음으로는 피부암 위치에서의 흡수선량률을 구했다. 피부암 위치에서의 흡수선량률을 알기 위해서는 피부 표면에서의 흡수선량률과 깊이에 따른 선량 백분율을 알아야 한다. 정확한 흡수선량률을 구하기 위해 저에너지 광자의 선량측정법을 다룬 TG-61 protocol에 나온 방법을 이용했다. 특성들이 측정된 후, 세포실험과 동물실험을 통해 치료 효과를 검증했다. 실험 결과 개발된 피상 전자 근접치료장비가 실제 치료에 적용할 수 있을 것이라고 판단했다. 또한, 상용화된 선형가속기와 비슷한 결과를 얻음으로써 추후 사업화 시에 유용한 점이 될 것으로 판단했다.