A localized volume selection technique using radial k-space scanning method and its application to $^{31}P$ spectroscopy are presented. This technique employes an additional radial gradient coil and double tuned RF coil for the effective volume localization. The advantages of the proposed localization technique are that the signal degradation due to $T_2$ decay can be reduced using short RF pulse and the gradient echo data acquisition scheme, and the volume selection is immune to chemical shift. The theoretical background of the technique, computer simulation, and experimental results obtained from human volunteer are provided. The proposed technique appears to provide the accurate volume localization which is especially useful to phosphorus localized volume spectroscopy with low SNR. All experiments have been performed in KAIS 2.0T superconducting NMR imaging system.
본 논문에서 공간주파수 영역의 방사형 주사 기법을 이용한 3-차원의 국부체적 선택 기법과 선택된 3-차원 국부체적 에서의 인 분광에 대한 응용을 제안하였다. 제안된 기법을 실험에 적용하기 위해서는 기존의 x-, y-, z- 방향의 경사자계에 새로운 방사형 경사자계를 추가해 사용 하였으며, 또한 인 분광에 대한 응용을 위해서 이중동조 RF 코일을 제작 하였다.
제안된 기법은 두개의 선택적 경사자계와 하나의 RF 펄스의 조합에 의해 3-차원의 국부체적을 선택 함으로써 에코타임을 매우 작게 줄일수 있으며, $T_1$ 이나 $T_2$ 이완에 의한 신호감소를 줄일수 있다. 실험 결과로써 선택된 3-차원 국부체적에 대한 영상과 고 분해능의 인 분광을 제시하였다. 본 논문에서 제안된 3-차원 국부체적 선택 기법은 여러 종류의 스펙트럼 편집 기법과 조합하여 사용 될수 있으며, 짧은 $T_2$ 를 갖는 성분에 대한 국부체적 분광법에 이용 될수 있을것으로 기대 된다.