We proposed a novel technique to obtain a fast ring-down pulses train by using only slow devices. The proposed system is a double-loop Mach-Zehnder interferometer for optaining continuous ring-down interferograms with two fiber-loop cavities having slightly different optical path lengths. Each group of pulses passing through the sample and reference loops creates each ring-down pulse train with equal time intervals in Vernier fashion, and produces multiple ring-down interferograms successively by interfering with each other with scanning of a delay line. The system requires a scanning range of only a few millimeters to obtain multiple ring-down interferograms. In a proof-of-concept demonstration, the intrinsic losses of two loops are estimated. The measured-loss of both loops is compared to the separately measured results with a conventional fiber-loop ring-down system.
저속 검출 장비만을 이용하여 고속 감쇠 신호를 획득할 수 있는 기술을 새롭게 제안하였다. 제안된 시스템은 약간의 광경로 차이를 갖는 두 개의 광섬유 루프 공진기로 구성된 연속 감쇠 간섭무늬 획득 가능한 이중 루프 마하젠더 간섭계이다. 감지 루프와 기준 루프를 통과하고 나온 각 펄스열은 각각 동일한 시간 간격을 갖는 버니어 형태의 감쇠 펄스열을 발생시키게 되며, 시간 지연을 통해 두 펄스열을 간섭시킴으로써 다중 감쇠 간섭무늬를 발생시킨다. 획득한 감쇠 간섭무늬를 통해 광섬유 루프의 손실을 측정해 내었다. 특히 본 시스템에서는 수 밀리미터의 주사 범위만을 이용하여 다중 감쇠 간섭무늬 획득이 가능하였다. 제안된 시스템을 검증하기 위해 일반적인 광섬유 감쇠 시스템을 이용하여 두 루프의 내부 손실을 각각 측정하였으며, 두 실험 결과를 서로 비교하였다.