원추형 끼워맞춤된 인공고관절의 축대칭 접촉 응력 해석 = Axisymmetric contact analysis of an artificial hip joint of the conical fitting type
서명 / 저자 원추형 끼워맞춤된 인공고관절의 축대칭 접촉 응력 해석 = Axisymmetric contact analysis of an artificial hip joint of the conical fitting type / 강영훈.
발행사항 [서울 : 한국과학기술원, 1987].
Online Access 제한공개(로그인 후 원문보기 가능)원문

소장정보

등록번호

4104715

소장위치/청구기호

학술문화관(문화관) 보존서고

MME 8751

휴대폰 전송

도서상태

이용가능(대출불가)

사유안내

반납예정일

리뷰정보

초록정보

The purpose of this study was to analyze the contact stress of the artificial hip joint of conical fitting type being subjected to axial force and frictional force on the contact surface. Frictional force was such that sliding was just to stop. Computed results were compared with the ones by ANSYS ,which turned out to be favorable. Some examples were studied to know the effect of cone angle and the effect of its misfitting on contact stresses. The results showed that stresses from the analysis with friction are smaller than those from the one without friction, and large cone angle produced small stresses.

서지기타정보

서지기타정보
청구기호 {MME 8751
형태사항 v, 60 p. : 삽화 ; 26 cm
언어 한국어
일반주기 부록 수록
저자명의 영문표기 : Young-Hoon Kang
지도교수의 한글표기 : 윤용산
지도교수의 영문표기 : Yong-San Yoon
학위논문 학위논문(석사) - 한국과학기술원 : 기계공학과,
서지주기 참고문헌 : p. 31-32
주제 Strains and stresses.
Friction.
Joints (Engineering)
인공 관절. --과학기술용어시소러스
접촉 응력. --과학기술용어시소러스
마찰력. --과학기술용어시소러스
맞춤 (설계) --과학기술용어시소러스
응력 해석. --과학기술용어시소러스
Artificial hip joints.