The aim of this study was to measure blood glucose by monitoring current generated from the reaction of glucose oxidase and glucose while glucose oxidase and mediator were immobilized on electrode formed with carbon paste. Screen printing was done to carbon paste to form electrode in which charge transfer reaction should occur, and strip sensor was produced by forming standard electrode using Ag/AgCl ink. On the surface of working electrode, enzyme membrane solution was formed using glutaraldehyde, ferrocene, hexammine ruthenium(Ⅲ) chloride, and glucose oxidase. Then enzyme membrane solution was immobilized. Using the strip glucose sensor, the amount of glucose in phosphate buffered saline (PBS) solution of pH 7.4, which is similar to human body, was measured using cyclic voltammetry and linear sweep method.
본 연구에서는 글루코스산화효소(glucose oxidase, GOx)와 전자전달매개체(mediator)를 카본페이스트로 형성시킨 전극위에 고정화하여, 글루코스산화효소가 글루코스와 반응하여 생성된 전자가 만들어내는 전류를 측정하여 글루코스의 농도를 측정하고자 하였다. 전자전달반응이 일어나야하는 working 전극은 카본 페이스트(C-992 (주)케티)를 이용하여 스크린프린팅을 통해 제작하였고, counter 전극도 카본 페이스트(C-992 (주)케티)를 통해 형성시켰다. Ag/AgCl ink를 이용하여 기준전극을 형성시켜 스트립형태의 센서를 제작하였다. 이렇게 만들어진 working electrode 전극 표면에 글루타르알데하이드(glutaraldehyde), 페로센(Ferrocene), 염화헥사아민루테늄(hexammine ruthenium chloride)과 글루코스산화효소를 이용하여 효소막용액을 만들어, 효소막을 고정화하였다. 이렇게 제작된 스트립형태의 글루코스 센서를 이용하여 생체조건과 유사한 pH 7.4 phosphate buffered saline(PBS)용액 내의 글루코스 양을 cyclic voltammetry와 linear sweep voltammetry의 방법을 통해 측정하였다.